成人VR视频

Quanta 450 ESEM 鈥 ESEM Operating Manual

Quanta 450 FE-SEM - ESEM Operating Manual (Revised with Safety Warnings & Tips)

Important: You must arrange with FEMR staff before operating the microscope in ESEM mode with the cooling (Peltier) stage.

Safety Warnings for the Quanta 450 ESEM
鈿狅笍 High鈥慥oltage and Electrical Safety
  • The microscope operates at high voltage鈥攏ever touch internal electrical components, detector contacts, or cables unless instructed by FEMR staff.
  • Do NOT plug or unplug detectors with the beam on or while the system is actively pumping.
  • Ensure all cables are fully seated; loose connections may cause arcing or damage to the detector.
鈿狅笍 Vacuum and Pressure Safety
  • Never open the chamber door until the system reads VENTED; pressurized gas release can cause injury or damage.
  • Do not obstruct or cover vacuum ports, vent lines, or pumping pathways.
  • Sudden pressure changes can cause sample material to eject鈥攕ecure samples properly.
鈿狅笍 Moisture and Condensation Hazards
  • Moisture inside the chamber is very dangerous for detectors and the pole piece.
  • Never exceed 80% humidity, and never bring the working distance closer than 8鈥痬m.
  • Be cautious about condensation and water leaks, as they can cause serious damage to the instrument.
  • If you see condensation forming on the stage, detectors, or chamber walls, stop immediately and call FEMR staff.
  • Do not insert visibly wet or dripping samples.
鈿狅笍 Cooling and Water System Safety
  • Never exceed the recommended minimum working distance (8鈥痬m)鈥攎oisture can enter the pole piece, causing permanent damage.
  • Use only distilled water in the chiller system鈥攐ther water types can corrode components.
  • Always check for water leaks after connecting the water lines. Stop immediately if leaks are detected.
  • Never run the Peltier stage without the water chiller switched on.
  • The cooling lines and stage may become very cold鈥攁void direct skin contact.
鈿狅笍 Sample Safety
  • Biological samples must be handled using PPE (gloves, lab coat, eye protection).
  • Unfixed biological material can contaminate the chamber鈥攃lean thoroughly with 70% ethanol afterward.
  • Do not load samples that exceed the height of the sample holder; they may contact the pole piece during pumping.
鈿狅笍 Detector Handling
  • Do not touch internal components (detectors, pole piece, stage hardware) unless explicitly instructed.
  • Handle detectors gently 鈥 Detectors (SED, GSED, GBSD) are fragile (and expensive)鈥攏ever pull on cables, connectors, or sensor heads.
  • Place removed detectors only on padded, clean surfaces.
鈿狅笍 Stage Movement Hazards
  • Keep hands clear of the stage and column when moving the stage.
  • Never raise the stage while looking away from the chamber image鈥攃rashes can permanently misalign the column.
  • When switching from the Peltier stage back to a standard holder, ensure the stage is fully lowered before inserting any hardware.
鈿狅笍 Chemical and Cleaning Safety
  • Only use approved cleaning agents (70% ethanol, lens tissue where appropriate).
  • Do not use acetone or strong solvents inside the chamber鈥攖hey can damage seals and coatings.
鈿狅笍 Thermal Safety
  • Cooled components can cause frostbite鈥憀ike burns鈥攁void direct contact.
  • Heated samples (if used) can cause burns鈥攗se tools, not fingers.
鈿狅笍 Emergency Shutdown Conditions

Immediately stop imaging and call FEMR staff if you observe:

  • Unusual noises from pumps or chiller
  • Water or condensation near the pole piece
  • Sudden pressure spikes or failure to reach vacuum
  • Smell of burning, ozone, or overheated components
  • Detector error warnings during ESEM operation


1. Prepare the Caron Water Chiller

  • Switch on the Caron water chiller behind the Quanta 450.
  • Confirm the temperature setpoint is 10鈥癈.
  • Fill reservoir with distilled water only (鈿狅笍 never deionized).

Safety Tip:
馃憠 If the chiller tank runs dry, the Peltier stage can overheat鈥攁lways check water level before activation.


2. Configure Initial Microscope Settings

  • Set Accelerating Voltage to 15 kV and Spot Size to 3.5.
  • Vent the chamber and open the door.

Safety Warning:
鈿狅笍 Wait for complete venting before opening the chamber door to avoid pressure鈥憆elated hazards.


3. Install the Peltier Cooling Stage

  • Remove the decelerating stage if installed.
  • Position the Peltier stage with water lines and ribbon cable facing backward.
  • Lightly tighten the helicoil center set screw.

Safety Tip:
馃憠 Do not overtighten the helicoil screw鈥攊t can strip easily.


4. Connect Cooling and Control Lines (Staff only)

  • Remove the water plugs inside the chamber.
  • Insert the water lines and connect the vertical 9鈥憄in ribbon cable.
  • Verify that the COOLING button activates.

Safety Warning:
鈿狅笍 Ensure the sample fits entirely within the holder鈥攑rotruding samples may contact the pole piece during pumping.


5. Prime the Water Valve System

  • Power on the Water Valve Box (left of column).
  • An alarm will sound鈥攑ress & hold the GREEN button until water flows smoothly with no air bubbles.
  • Check for any leaks.

Critical Warning:
鈿狅笍 Immediately stop if water leaks inside the chamber鈥攁lert staff before continuing.


6. Install the Appropriate Detector (staff only)

  • Carefully remove the Wide鈥慉ngle SED.
  • Install the GSED or GBSD and clip the axial endpiece to the pole piece.
  • Close the chamber and press VACUUM.

Safety Tip:
馃憠 Hold detectors by their housings鈥攏ot their cables or sensor heads.


7. Cooling & Humidity Control

  • Activate cooling under the Temperature Control tab.
  • Open the Humidity tab to view the Triple鈥慞oint Graph.
  • Recommended starting point:
    • Temp: 5鈥癈
    • Pressure: 5.5 Torr
      鈫 Produces ~75鈥78% humidity (stable, moderate moisture)

Safety Warning:
鈿狅笍 High humidity (>80%) can cause water droplets to form on internal components鈥攔isking electrical and vacuum damage.


8. Adjust Temperature & Pressure

  • Input desired Pressure in the main menu.
  • Adjust Temperature 鈫 press GO TO.
  • Monitor stabilization under the Humidity tab.

Tip:
馃憠 Make small adjustments鈥擡SEM humidity is highly sensitive and changes rapidly.


9. Pre鈥慖maging Setup

  • Press BEAM ON.
  • Center the stage.
  • Raise stage until Peltier stage edge is just above the yellow 10鈥痬m line.

Critical Warning:
鈿狅笍 Never approach closer than 8鈥痬m to the pole piece鈥攎oisture can be pulled into the column.

  • At 100脳, increase to 400脳 and use Auto Brightness/Contrast.
  • Focus on the bottom of the sample holder.

Safety Tip:
馃憠 Lens alignment should be done now鈥攊t is difficult in high鈥慼umidity ESEM mode.


10. Load the Sample

  • Turn BEAM OFF.
  • Vent the chamber fully.
  • Pre鈥慶ool the sample when possible.
  • Load onto stage; optionally add water to the four dimples.
  • Close chamber and press PUMP.

Biological Safety Warning:
鈿狅笍 Wear gloves and avoid introducing contamination inside the chamber.


11. Begin Imaging

  • Press BEAM ON and start at 400脳.
  • Find a feature to focus on.
  • Link stage height.
  • Set WD 鈮 8.00 mm.

Safety Tip:
馃憠 Do not adjust WD below 8.00 mm even if the sample appears flat鈥攔isk of condensation on the pole piece increases at close distances.


12. End-of-Session Procedures

  • Restore all settings to defaults:
    • Cooling Off
    • Pref:Scan:ESEM default
    • Stage lowered fully
    • WD fully down
  • Drain water lines; disconnect water and control cables.
  • Remove the Peltier stage, reinstall the single-stub holder, and reinstall the SED (Staff only).

Safety Warning:
鈿狅笍 Water remaining in lines can backflow into the microscope鈥攅nsure they are fully drained before disconnecting.


13. Cleaning & Decontamination

  • Clean all holders, counters, and nearby surfaces.
  • Remove all particles, tubes, or biological debris.
  • After unfixed biological samples, wipe with 70% ethanol.

Tip:
馃憠 Allow surfaces to air-dry fully before closing the chamber to avoid moisture spikes during the next session.


14. Final Safety Note

Before inserting the standard specimen holder, always fully lower the stage after removing the Peltier stage to prevent collision with the pole piece.

Back to top